The key distinction is not whether software is used in healthcare, but whether it performs a medical function that brings it within the medical device definition. A patient portal, appointment scheduling platform, and AI diagnostic tool may all be part of the same healthcare ecosystem, yet they do not necessarily receive the same regulatory treatment.
The distinction matters as clinical software becomes more common. The American Medical Association reported that 66% of surveyed physicians used healthcare AI in 2024, including tools for documentation, clinical decision support, and assistive diagnosis. As these capabilities move into healthcare products, software teams need to understand when a feature crosses from administrative support into a regulated medical function.
This guide explains how to assess software as a medical device (SaMD) under the U.S. regulatory framework, distinguish regulated functions from general healthcare features, and understand the development responsibilities that may follow.
Key Takeaways
- Software qualifies as SaMD when it performs an independent medical purpose, such as diagnosing, monitoring, or treating a disease.
- Regulatory status depends on the software’s intended use and specific functions, not its platform, programming language, or use of AI.
- Administrative healthcare software and certain clinical decision support functions are excluded from the medical device definition.
- A function may meet the device definition but receive a different level of regulatory oversight based on its risk.
- SaMD development requires documented intended use, risk management, software validation, and appropriate regulatory planning.
What Is Software as a Medical Device (SaMD)?
Software as a Medical Device (SaMD) is software intended to perform one or more medical purposes without being part of a hardware medical device. It can operate on general-purpose platforms, including smartphones, computers, cloud environments, and servers.
The International Medical Device Regulators Forum (IMDRF) established this definition to distinguish standalone medical software from software that operates as part of a physical medical device.
Examples include:
- Software that analyzes medical images to identify potential abnormalities.
- Applications that interpret cardiac signals to detect irregular heart rhythms.
- Algorithms that calculate a patient’s risk of developing a specific disease.
- Software that analyzes patient data to support treatment planning.
- Remote monitoring applications that interpret physiological measurements and identify clinically significant changes.
SaMD does not need to run on specialized medical hardware. A mobile application or cloud-based platform can qualify if its intended purpose and functionality meet the medical device definition.
When Does Software Qualify as a Medical Device Under FDA Regulations?

Healthcare software qualifies as a medical device when its intended purpose and functionality meet the applicable medical device definition under the Federal Food, Drug, and Cosmetic Act (FD&C Act).
The assessment depends on what the software is designed and marketed to do, how its functions operate, and the medical decisions or outcomes they are intended to support.
1. Intended Medical Purpose
The first consideration is whether the software is intended to diagnose, prevent, monitor, or treat a disease or medical condition, or perform another recognized medical purpose.
For example, an application that records a patient’s blood pressure is not automatically a medical device. However, software specifically designed to analyze those readings, identify signs of hypertension, and provide a diagnostic assessment may qualify.
Product documentation, labeling, marketing claims, and the software’s actual design can all help establish its intended purpose.
2. Actual Software Functionality
The next step is to examine what each function does with the information it receives.
Consider a healthcare platform with these capabilities:
- Patient registration and appointment scheduling.
- Electronic health record storage.
- Display of laboratory results.
- Analysis of medical images.
- Prediction of disease risk based on patient data.
The first three functions generally support administrative or information-management activities. The image analysis and disease-risk prediction functions may qualify as medical device functions because they interpret clinical information for a medical purpose.
This distinction is important because a single application can contain both regulated and non-regulated functions. The FDA’s software policy guidance explains how regulatory oversight applies to device software functions, including those running on mobile and general-purpose computing platforms.
The practical rule: assess each software function according to its intended use and clinical role. Do not classify an entire healthcare platform based only on its product category or marketing description.
Difference Between SaMD, SiMD, and General Healthcare Software
SaMD, Software in a Medical Device (SiMD), and general healthcare software serve different purposes. Their main distinction is whether the software performs an independent medical function, operates as part of medical hardware, or supports healthcare administration and information management.
| Software Category | What it Does | Example |
| Software as a Medical Device (SaMD) | Performs an independent medical purpose without being part of medical hardware | Standalone software that analyzes ECG data to detect cardiac abnormalities |
| Software in a Medical Device (SiMD) | Operates within or controls a physical medical device to achieve its intended purpose | Embedded software controlling an infusion pump |
| General healthcare software | Supports administrative, communication, or information-management activities without performing a medical device function | Appointment scheduling or patient registration software |
The distinction between SaMD and SiMD depends on the relationship between the software and physical medical equipment. Software that controls a medical device or is necessary for its hardware to achieve its intended medical purpose is generally treated as software in a medical device rather than standalone SaMD.
General healthcare software falls outside the medical device category when its functions do not meet the applicable device definition or qualify for a statutory exclusion.
Which Healthcare Software Functions Are Not Considered Medical Devices?
Not every application used by hospitals, clinics, or patients falls under medical device regulation. U.S. law specifically excludes certain software functions from the device definition, including administrative support, some electronic records, data storage and transfer, and qualifying clinical decision support.
The exclusions are set out in Section 520(o) of the Federal Food, Drug, and Cosmetic Act.
| Software function | Typical Example | Regulatory consideration |
| Administrative support | Billing, appointment scheduling, claims processing, and inventory management | Excluded when limited to the specified administrative purposes |
| Electronic patient records | Software that stores and displays patient-provided information | May be excluded when it meets the statutory conditions |
| Data transfer and storage | Software that transfers, stores, converts, or displays medical information | Excluded when it does not interpret or analyze the data for a medical purpose |
| General wellness | Applications that track activity, sleep, or general fitness | May be excluded when unrelated to diagnosing, preventing, or treating disease |
| Clinical decision support | Software that provides evidence-based information to help clinicians assess treatment options | May qualify for the statutory non-device CDS exclusion if all required criteria are met |
These exclusions depend on the actual functionality. For example, software that displays laboratory results without interpreting them may fall outside the device definition. Software that analyzes those results to identify a disease or recommend a specific treatment may require a different assessment.
A separate distinction is enforcement discretion. A software function may meet the medical device definition, but the FDA may decide not to enforce certain requirements against it because of its risk profile. That does not mean the function is legally excluded from the device definition.
What Are Examples of Software as a Medical Device?
Software qualifies as SaMD when it independently performs a medical function rather than simply storing, displaying, or organizing healthcare information.
Common examples include:
| SaMD application | Medical function |
| Medical imaging analysis | Analyzes CT, MRI, or X-ray images to identify potential abnormalities |
| ECG analysis software | Interprets cardiac signals to detect possible arrhythmias |
| Disease prediction software | Uses patient information to estimate the likelihood of a specific disease |
| Clinical monitoring software | Analyzes physiological data to identify clinically significant changes |
| Treatment planning software | Calculates or recommends patient-specific treatment parameters |
| Diabetes management software | Analyzes glucose measurements to support disease management or treatment decisions |
These examples are not automatic regulatory classifications. The software’s intended use, claims, users, inputs, outputs, and clinical role determine whether a particular function meets the medical device definition.
For instance, a mobile application that records glucose readings is not necessarily SaMD. An application that analyzes those readings to recommend insulin doses performs a substantially different function and requires a closer regulatory assessment.
When Is Clinical Decision Support Software Regulated by FDA?

Clinical Decision Support (CDS) software may be regulated as a medical device when it does not meet the statutory conditions for exclusion from the device definition.
The 21st Century Cures Act created a specific exclusion for certain clinical decision support functions. To qualify, a software function must meet all four statutory criteria:
- It must not acquire, process, or analyze certain medical images or signals. The exclusion does not cover software functions that process specified medical images or signals acquired from in-vitro diagnostic devices or signal-acquisition systems.
- It must operate on medical information. The function must display, analyze, or print medical information about a patient or other medical information.
- It must support recommendations to healthcare professionals. The software must provide recommendations concerning the prevention, diagnosis, or treatment of a disease or condition.
- It must allow independent professional review. Healthcare professionals must be able to independently review the basis for the recommendations rather than primarily relying on the software’s output.
If a function fails to meet any one of these criteria, it does not qualify for this particular non-device CDS exclusion. It may still be outside the device definition for another reason, or it may qualify as a device function.
For example, software that presents relevant clinical guidelines and patient information for a physician to evaluate may qualify for the exclusion. In contrast, software that provides a specific diagnostic result, disease-risk score, or treatment directive may fall outside it.
Patient-facing diagnostic applications also require careful assessment. The statutory CDS exclusion is designed around recommendations to healthcare professionals, not software that independently provides diagnostic or treatment recommendations to patients.
How Does FDA Classify and Regulate SaMD?
Once a software function qualifies as a medical device, its regulatory classification depends on the level of control necessary to provide reasonable assurance of safety and effectiveness.
The U.S. system uses three medical device classes:
| FDA Device Class | Risk Level | General Regulatory Approach |
| Class I | Low risk | General controls, with some devices exempt from certain premarket requirements |
| Class II | Moderate risk | General controls and special controls; many devices require 510(k) clearance |
| Class III | High risk | Generally requires premarket approval (PMA) |
The classification determines which regulatory controls and marketing pathway apply. It does not mean every software product must follow the same approval process.
Common FDA Marketing Pathways
- 510(k) Premarket Notification: Manufacturers demonstrate that their device is substantially equivalent to a legally marketed predicate device. This pathway is common for eligible Class II devices and some Class I or Class III devices.
- De Novo Classification: A pathway for novel devices that meet the device definition but are low to moderate risk and lack a suitable predicate device. A De Novo request can establish a new classification and create a predicate for future devices.
- Premarket Approval (PMA): A more rigorous pathway generally used for Class III devices, requiring evidence supporting the device’s safety and effectiveness.
Some devices are exempt from premarket notification, subject to applicable limitations. Others may require registration, listing, quality system compliance, labeling controls, and postmarket obligations depending on their classification and regulatory status.
Manufacturers should identify the applicable device type, product code, classification, and pathway before planning a market launch. The FDA’s medical device classification guidance is a useful starting point.
One important distinction: IMDRF’s SaMD risk framework and the FDA’s Class I, II, and III system are not interchangeable. IMDRF categories consider the significance of the healthcare situation and the software’s impact on a healthcare decision. FDA classification follows its own device classification and regulatory-control framework.
What Should Developers Consider When Building SaMD?

SaMD development requires more than building a functional application. Teams must account for medical intended use, patient safety, software reliability, clinical evidence, and regulatory obligations throughout the product lifecycle.
1. Define Intended Use and Product Claims
Document the medical purpose, intended users, target patient population, clinical environment, and expected output. Product claims should match the functions the software actually performs.
Unclear intended-use statements can create regulatory uncertainty, particularly when a platform combines administrative features with diagnostic or treatment-related capabilities.
2. Design for Risk Management
Identify hazards that could result from incorrect calculations, inaccurate predictions, delayed alerts, data corruption, or software failure.
Assess the likelihood and potential severity of harm, then define controls to reduce those risks. The ISO 14971 medical device risk management standard provides a framework for identifying, evaluating, controlling, and monitoring risks throughout a device’s lifecycle.
3. Establish Software Lifecycle Controls
Medical device software needs documented development processes, requirements traceability, version control, change management, and verification activities.
IEC 62304 defines a framework for medical device software lifecycle processes, including development and maintenance activities. It supports a structured engineering process, but following the standard alone does not establish regulatory compliance.
4. Validate Clinical Performance
Software verification checks whether the product was built according to its specifications. Validation evaluates whether it performs as intended for its actual users and use environment.
For SaMD, this may involve testing algorithms against representative clinical datasets, assessing false positives and false negatives, evaluating usability, and establishing evidence that supports the intended medical claims.
AI-enabled software may require additional evaluation for model performance, bias, data quality, and changes in performance after deployment.
5. Address Cybersecurity and Data Protection
Security failures can affect software availability, clinical data integrity, and patient safety. Development teams should consider access controls, encryption, secure APIs, vulnerability management, audit logging, and incident response.
These controls should be incorporated into the product architecture rather than added after development.
6. Plan for Maintenance and Postmarket Monitoring
SaMD responsibilities do not end when the software launches. Teams may need to monitor complaints, investigate performance issues, manage updates, address vulnerabilities, and assess whether product changes affect its safety or regulatory status.
For AI-enabled medical software, changes to algorithms or training data can affect clinical performance. A planned change process helps teams evaluate those changes before deployment.
How Can You Determine Whether Your Software Is SaMD?
Use a function-level assessment before finalizing the product architecture or development scope. The following questions provide a practical starting point:
- What medical purpose does the software serve? Identify whether it is intended to diagnose, prevent, monitor, or treat a disease or perform another medical purpose.
- What does each function do? Separate administrative features, information storage, data display, and clinical analysis.
- Does any function interpret patient data for a medical purpose? Review diagnostic outputs, risk scores, treatment recommendations, and clinical alerts.
- Does a statutory exclusion apply? Check whether the function meets the conditions for administrative, wellness, electronic-record, data-transfer, or non-device CDS exclusions.
- What regulatory pathway may apply? If a function meets the device definition, assess its classification, applicable controls, and potential marketing pathway.
Document the reasoning behind each decision. If the intended use or regulatory status remains unclear, seek advice from a qualified medical device regulatory professional or obtain appropriate regulatory feedback before commercialization.
For a broader view of the technical, security, and compliance decisions involved, consult the healthcare software development guide.
Conclusion
Software becomes a medical device when its intended purpose and functionality meet the applicable medical device definition. For healthcare technology teams, the critical task is to assess individual functions, understand statutory exclusions, and determine the regulatory controls that apply before bringing the product to market. A scheduling feature and a diagnostic algorithm may sit inside the same application, but they do not necessarily carry the same regulatory responsibilities.
Building SaMD calls for close coordination between product, engineering, clinical, quality, and regulatory teams. If your product includes clinical analysis, patient monitoring, or decision-support capabilities, Cubix healthcare software development services can help you plan the technical foundations around the product’s intended use, safety, and development requirements.
Planning a Medical Software Product? Start with the Right Team.
Contact Us
Frequently Asked Questions
1. Is All Healthcare Software Considered a Medical Device?
No. Administrative platforms, billing systems, patient portals, and certain record-management tools are generally outside the medical device definition when their functions do not perform a regulated medical purpose or qualify for an exclusion.
2. What Is an Example of Software as a Medical Device?
Standalone software that analyzes medical images, interprets ECG signals, or calculates patient-specific disease risk can qualify as SaMD when intended for a medical purpose.
3. Is an Electronic Health Record (EHR) System a Medical Device?
An EHR system used to store, transfer, and display patient records generally does not qualify as a medical device when it meets the statutory conditions. Individual functions that analyze clinical information may require a separate assessment.
4. Does AI Healthcare Software Require FDA Approval?
Not automatically. AI software requires regulatory assessment when its intended use and functionality meet the medical device definition. The applicable requirements depend on the specific function, device classification, and marketing pathway.
5. What Is the Difference Between SaMD and SiMD?
SaMD performs an independent medical purpose without being part of medical hardware. SiMD is software that operates within or controls a physical medical device to achieve its intended purpose.